Key result
The 56-thread Multilayer Flow Modulator reduces stenosis ~49% versus 80-96 thread devices.
Why the study?
The impact of design thread count and duration of implantation on the biocompatibility of the Multilayer Flow Modulator in porcine models was not well characterized.
Does the 56-thread Multilayer Flow Modulator improve biocompatibility and reduce stenosis compared to the 80-96 thread design in a porcine model?
Does the 56-thread Multilayer Flow Modulator improve biocompatibility and reduce stenosis compared to the 80-96 thread design in a porcine model?
The 56-thread Multilayer Flow Modulator device elicited less intra-arterial response and lower percentage stenosis compared to the 80-96 thread device in a porcine model.
Lower-thread MFM devices induced less neointimal hyperplasia in pigs; leaves open optimal design selection for human aneurysm trials.
Background: The Multilayer Flow Modulator (MFM) (Cardiatis, Isnes, Belgium) is a self-expandable mesh of cobalt alloy wires used for the treatment of aortic aneurysms.The impact of design thread count and duration of implantation on the biocompatibility of the MFM in porcine animal models was assessed in this study.Methods: Eight mini-piglets received 26 MFM devices (12 with 56 threads, 14 with 80-96 threads) in the iliac, carotid, and renal arteries.Animals were sacrificed and specimens explanted at 1, 3, and 6 months, at which time histological and ultrastructural analyses were carried out.Results: The MFM was successfully deployed in 25 of the 26 implanted cases.The 56 thread devices were well tolerated locally and yielded fewer signs of inflammation and neo intimal hyperplasia.Percentage stenosis was 16.9% ± 5.1% for the 56 thread devices versus 33.4% ± 10.2% for the 80-96 thread devices (p=0.001) at 3 months, and 21.7% ± 9.9% for the 56 thread devices versus 33.6% ± 12.4% for the 80-96 thread devices (p=0.004) at 6 months.The 5 devices selected for SEM examination were well deployed, integrated into the vessel wall and endothelialized, and had patent side branches.Conclusions: No significant stenosis formation or inflammatory response was recorded in any of the implanted animals.The 80-96 thread devices elicited a greater intra-arterial response than the 56 thread devices, although the values for both groups remained within the normal range for stented carotid, renal, or iliac arteries.Further preclinical and clinical studies will extend assessment of the long-term safety and effectiveness of the MFM.
No takes yet. Share an insight, caveat, or question.
A 2014 study studied Healthy arteries (assessment for aortic aneurysm treatment) (n=8). Multilayer Flow Modulator (56 threads) vs. Multilayer Flow Modulator (80-96 threads) was evaluated on Percentage stenosis at 3 months (p=0.001). The 56-thread Multilayer Flow Modulator devices yielded significantly lower percentage stenosis compared to the 80-96 thread devices at 3 months (16.9% vs 33.4%, p=0.001) and 6 months (21.7% vs 33.6%, p=0.004).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: