We compared reticulocyte levels in 85 patients determined by conventional manual methods and two automated methods, image analysis on the Hematrak 590 and flow cytometry on the Ortho Spectrum. Each automated method showed acceptable correlation with manual determinations over a range of 0.4% to 31% reticulocytes. The Hematrak 590 regression equation was y=0.98x+0.37. The Ortho Spectrum regression equation was y=0.95x+0.87. Automated determination of reticulocyte percentage by image analysis or flow cytometry is an acceptable alternative to manual counts, especially for high-volume laboratories. The advantages of each automated methodology are compared.