The central electromagnetic calorimeter for the Collider Detector at Fermilab was calibrated in a test beam of 50 GeV/ c electrons. This calibration is maintained on long time scales with individual 137 Cs sources for each module, and on short time scales with individual LED and xenon flasher systems on each module. The ratio of the 50 GeV/ c electron response to the response to a 137 Cs source is assumed to be constant; tests on the order of a month indicate a reproducibility of ± 0.4%. An average 2% loss of light yield over a period of 2 years has been observed.