Key result
Simulated RFA ablation volume varies with blood flow and tissue deformation at ≥1.5 mm depth.
Why the study?
Computational modeling of radio frequency cardiac ablation can provide efficient and reliable estimates of ablation volume and treatment outcomes, improving upon ex vivo and in vivo studies.
A fully coupled electrothermomechanical computational model demonstrates that predicted RFA ablation volume depends significantly on blood flow rate and tissue deformation from electrode insertion depth.
RFA planning models must account for perfusion and deformation; hypothesis-generating for patient-specific ablation until validated in vivo.
The application of radio frequency ablation (RFA) has been widely explored in treating various types of cardiac arrhythmias. Computational modeling provides a safe and viable alternative to ex vivo and in vivo experimental studies for quantifying the effects of different variables efficiently and reliably, apart from providing a priori estimates of the ablation volume attained during cardiac ablation procedures. In this contribution, we report a fully coupled electrothermomechanical model for a more accurate prediction of the treatment outcomes during the radio frequency cardiac ablation. A numerical model comprising of cardiac tissue and the cardiac chamber has been developed in which an electrode has been inserted perpendicular to the cardiac tissue to simulate actual clinical procedures. Temperature-dependent heat capacity, electrical and thermal conductivities, and blood perfusion rate have been considered to model more realistic scenarios. The effects of blood flow and contact force of the electrode tip on the treatment outcomes of a fully coupled model of RFA have been systematically investigated. The numerical study demonstrates that the predicted ablation volume of RFA is significantly dependent on the blood flow rate in the cardiac chamber and also on the tissue deformation induced due to electrode insertion depth of 1.5 mm or higher.
No takes yet. Share an insight, caveat, or question.
Singh et al. (2020) studied Cardiac arrhythmias. Radio frequency ablation (RFA) computational model was evaluated on Ablation volume. A fully coupled electrothermomechanical model demonstrated that predicted ablation volume of RFA is significantly dependent on blood flow rate and tissue deformation from electrode insertion depth ≥1.5 mm.