Key result
Passive fixation atrial lead spontaneously dislodges 9 years post-implant in rare autopsy case.
Why the study?
Atrial lead dislodgement typically occurs within 2 months after implantation, but extremely late migration over 5 years is rare and poorly understood.
Case Report (n=1)
Highlights the rare possibility of extremely late atrial lead dislodgement occurring years after implantation.
Very late passive-fixation atrial lead dislodgement can occur without trigger; this case extends known timelines but leaves incidence and mechanisms open.
The complication rate of pacemaker insertion is 3.8%–13.2%, and the most frequent lead-related one is dislodgement, with 0.79% (procedure-year)1–4; this complication mostly occurs within 2 months after the procedure. Atrial lead migration >5 years after implantation is extremely rare, and little has been reported on further observation of the causes, such as pathologic anatomy or patient characteristics. Here, we describe a case in which an atrial lead with passive fixation dislodged 9 years after the implantation without an obvious trigger and in which an autopsy was performed for the pursuit of causes.
No takes yet. Share an insight, caveat, or question.
Nakamura et al. (2020) conducted a case report in Atrial lead dislodgement (n=1). Atrial lead with passive fixation was evaluated on Atrial lead dislodgement. An atrial lead with passive fixation dislodged 9 years after implantation without an obvious trigger, representing an extremely rare case of late migration evaluated via autopsy.