Key result
Photoelastic stress analysis accurately measures vascular stress with ~4% error across physiological pressures.
Why the study?
Real-time and accurate stress calculation in vasculature walls is needed to provide feedback control for catheter insertion robots without altering catheter properties.
A calibrated photoelastic stress analysis system achieved an average error of 3.9% in stress measurements in vasculature models, enabling accurate real-time feedback for catheter insertion robots.
Supports bench validation of stress-sensing catheter systems; leaves open in vivo translation before clinical use.
Real-time and accurate stress calculation in walls of vasculature is a desired goal in order to provide feedback control for catheter insertion robots without changing catheter stiffness and lumen. This feedback source also has applications in endovascular surgery simulation. In order to address this need, we consider photoelastic effect, as birefringence produced by light retardation relates to the stress inside photoelastic materials. In this research, a polariscope was designed for urethane elastomer vasculature models, the photoelastic coefficient of urethane elastomer was measured, and a camera system was calibrated to quantify and reduce error in the measurement system. An average error of 3.9% was found in stress measurements for the pressure range of 60-189 mmHg inside the urethane elastomer model. This result was applied to correct stress distribution images, using the average stress value as a reference and preserving local maxima and minima of stress. This enabled us to accurately calculate stress in vasculature models during human blood pressure simulation (HBPS), and enables the comparison, in a closed loop, of stress produced by HBPS and by catheter motion when driven by a robot, as well as to measure the stress produced by medical tools on the vascular model wall.
No takes yet. Share an insight, caveat, or question.
Tercero et al. (2010) studied this question. Photoelastic stress analysis was evaluated on Stress measurement error. Photoelastic stress analysis in urethane elastomer vasculature models demonstrated an average error of 3.9% for pressure ranges of 60-189 mmHg, enabling accurate stress calculation during simulation.