Key result
False troponin elevations from heterophile antibodies correlate strongly with hemoglobin levels in a case report.
Why the study?
Falsely elevated troponin values caused by assay interference have been reported, but fluctuation correlating with hemoglobin as a marker of heterophile antibody levels is not well described.
Case Report (n=1)
Hemoglobin trending may serve as a surrogate measure of heterophile antibody titers in cases of falsely elevated troponin.
Answers are grounded in this paper and cite their sources.
Alerts to troponin assay interference risk; hemoglobin correlation is hypothesis-generating and requires validation before clinical use.
Serum troponin is a useful laboratory study for the diagnosis of acute myocardial infarction. However, elevations can also be seen in a variety of other diseases processes. Falsely positive troponin values caused by interference with current troponin assays have been reported. We report a unique case that demonstrates the fluctuation of falsely elevated troponin correlating with hemoglobin, serving as a marker of heterophile antibody levels. A 74-year-old gentleman presented to our Emergency Department with a several-day history of increasing shortness of breath associated with a new-onset chest pain and a troponin I level of 77.28 ng/mL. Throughout his stay, fluctuations in measured troponin levels correlated strongly with fluctuations in hemoglobin levels. Several investigations confirmed false elevated troponin levels secondary to heterophile antibody interference. We conclude that hemoglobin trending in our patient represented a surrogate measure of his heterophile antibody titers with time and that fluctuations in these levels correlated with respective fluctuations in the falsely elevated troponin levels.
No takes yet. Share an insight, caveat, or question.
Ghali et al. (2012) conducted a case report in Falsely elevated troponin I (n=1). Heterophile antibody interference was evaluated on Fluctuation of falsely elevated troponin correlating with hemoglobin. Falsely elevated troponin levels secondary to heterophile antibody interference fluctuated in strong correlation with hemoglobin levels in a 74-year-old man.