Key result
Leadless Micra pacemaker shows near-complete endothelialization just 4 months post-implant in a single case.
Why the study?
The degree of endothelialization of leadless pacemaker systems and implications for explantation strategies are not well described.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Leadless pacemakers can become nearly completely endothelialized within 4 months of implantation, suggesting that future extraction may require cardiac surgery.
A 43-year-old man with congenital heart disease (patent ductus arteriosus and congenital aortic stenosis) underwent leadless pacemaker (Medtronic Micra) implantation in September 2016. In 2007, aortic valve replacement was complicated by complete A–V block requiring implantation of a DDD-pacemaker. Over the subsequent 3 years, he developed progressive heart failure and his course was complicated by recurrent device-related infections. In early 2015, he underwent placement of an epicardial pacing system which also became infected. With insufficient intrinsic rhythm and without traditional venous or epicardial access, we believed the only pacing-option for this patient was a leadless pacemaker as a bridge to transplant even though a VVIR system was not optimal. Therefore, in August 2015, the patient underwent uneventful implantation of a leadless pacemaker (See supplementary video 1 and supplementary figure S2 ) and 4 months thereafter underwent orthotopic heart transplant. In the explanted native heart, the leadless pacemaker was found to be embedded within the cardiac wall and nearly completely endothelialized. Two of the four pacemaker tines were embedded entirely within the cardiac wall and the remaining two were nearly covered by endothelial tissue (See Panel A and supplementary figure S1 ). We believe these findings may be generalizable for patients with leadless pacemaker systems and further inform strategies for explantation. While the risk of infection is lower with leadless pacemaker systems, we would hypothesize that extraction with this degree of endothelialization would likely require cardiac surgery. Invasive or non-invasive imaging strategies to assess the degree of endothelialization in patients with leadless pacemaker systems requiring explant may be of use. Panel A The explanted heart with the leadless pacemaker embedded in the right ventricular wall (arrow). The enlargement shows the pacemaker after it has been removed from the myocardium. Supplementary material is available at European Heart Journal online .
No takes yet. Share an insight, caveat, or question.
Borgquist et al. (2016) conducted a case report in Congenital heart disease with complete A-V block and recurrent device-related infections (n=1). Leadless pacemaker (Medtronic Micra) implantation was evaluated on Endothelialization of the leadless pacemaker in the explanted heart. In a 43-year-old man, a leadless Medtronic Micra pacemaker was found to be nearly completely endothelialized and embedded within the cardiac wall just 4 months after implantation.
Synapse has enriched 10 related papers. Consider them for comparative context: