A high performance liquid chromatography (HPLC) system complemented with T-split, capillary detection cell and a high voltage power supply was used for peptide mapping by gradient electrochromatography and nanoliquid chromatography (nano-LC). With capillary columns of 100 microm ID, 6 cm packed with octadecylated 1.5 microm silica particles, the typical analysis time was approximately 10-15 min. The resolution of a tryptic digest of cytochrome c obtained by electrochromatography at 100 kV/m was superior compared to the analysis by nano-LC. Bubble formation caused by Joule heating at currents up to 100 microA was successfully suppressed by using a resistor capillary of 25 microm ID connected to the outlet of the packed column.