Key result
Noninvasive StO2 demonstrates a moderate ~0.58 correlation with ScvO2 after pediatric cyanotic congenital heart surgery.
Why the study?
The correlation between noninvasive tissue oxygen saturation (StO2) and serum central venous oxygen saturation (ScvO2) in pediatric patients with congenital cyanotic heart disease after cardiac surgery was not established.
Does noninvasive tissue oxygen saturation monitoring correlate with central venous oxygen saturation in pediatric patients with postoperative congenital cyanotic heart disease?
Observational (n=13)
No
Does noninvasive tissue oxygen saturation monitoring correlate with central venous oxygen saturation in pediatric patients with postoperative congenital cyanotic heart disease?
Effect estimate: Pearson's correlation of .58 (95% CI .35-.75)
p-value: p=<.0001
Noninvasive tissue oxygen saturation monitoring of the buccal mucosa shows a moderate correlation with central venous oxygen saturation, suggesting it may be a useful adjunct for monitoring low cardiac output after pediatric cardiac surgery.
Using a novel noninvasive, visible-light optical diffusion oximeter (T-Stat VLS Tissue Oximeter; Spectros Corporation, Portola Valley, CA) to measure the tissue oxygen saturation (StO2) of the buccal mucosa, the correlation between StOz and central venous oxygen saturation (ScvO2) was examined in children with congenital cyanotic heart disease undergoing a cardiac surgical procedure. Paired StO2 and serum ScvO2 measurements were obtained postoperatively and statistically analyzed for agreement and association. Thirteen children (nine male) participated in the study (age range, 4 days to 18 months). Surgeries included Glenn shunt procedures, Norwood procedures, unifocalization procedures with Blalock-Taussig shunt placement, a Kawashima/ Glenn shunt procedure, a Blalock-Taussig shunt placement, and a modified Norwood procedure. A total of 45 paired StO2-ScvO2 measurements was obtained. Linear regression demonstrated a Pearson's correlation of .58 (95% confidence interval [CI], .35-.75; p < .0001). The regression slope coefficient estimate was .95 (95% CI, .54-1.36) with an interclass correlation coefficient of .48 (95% CI, .22-.68). Below a clinically relevant average ScvO2 value, a receiver operator characteristic analysis yielded an area under the curve of .78. Statistical methods to control for repeatedly measuring the same subjects produced similar results. This study shows a moderate relationship and agreement between StO2 and ScvO2 measurements in pediatric patients with a history of congenital cyanotic heart disease undergoing a cardiac surgical procedure. This real-time monitoring device can act as a valuable adjunct to standard noninvasive monitoring in which serum SyvO2 sampling currently assists in the diagnosis of low cardiac output after pediatric cardiac surgery.
No takes yet. Share an insight, caveat, or question.
Yadlapati et al. (2013) conducted an observational in Postoperative congenital cyanotic heart disease (n=13). Noninvasive tissue oxygen saturation (StO2) monitoring vs. Central venous oxygen saturation (ScvO2) monitoring was evaluated on Correlation between StO2 and ScvO2 (Pearson's correlation of .58, 95% CI .35-.75, p=<.0001). Noninvasive tissue oxygen saturation (StO2) demonstrated a moderate correlation (r = 0.58) with central venous oxygen saturation (ScvO2) in pediatric patients after cyanotic congenital heart surgery.