Key result
HIV infection is linked to ~2.6-fold higher rates of appropriate ICD therapies versus matched controls.
Why the study?
There is limited data on the outcomes of implantable cardioverter-defibrillators in HIV-infected individuals despite their increased risk of sudden cardiac death.
Does HIV infection increase the rate of appropriate ICD therapies in patients with implantable cardioverter-defibrillators compared to uninfected controls?
Cohort (n=71)
No
Does HIV infection increase the rate of appropriate ICD therapies in patients with implantable cardioverter-defibrillators compared to uninfected controls?
Answers are grounded in this paper and cite their sources.
Loading...
HIV-infected patients with ICDs experience significantly higher rates of appropriate and inappropriate therapies compared to matched uninfected controls, particularly those with detectable viral loads.
BACKGROUND: HIV-infected individuals have a known increased risk of sudden cardiac death (SCD) compared to uninfected individuals. Implantable cardioverter-defibrillators (ICDs) are standard therapy for preventing SCD; however, there is limited data on the outcomes of ICDs in HIV-infected individuals. HYPOTHESIS: HIV-infected subjects receive a higher number of appropriate ICD therapies than uninfected controls. METHODS: This is a retrospective cohort study of 35 consecutive HIV-Infected patients and 36 uninfected controls matched by age, race, and gender who were treated at the University of North Carolina Medical Center in the outpatient or inpatient setting from 2014 to the present and had undergone ICD implantation. For HIV-infected subjects, a multivariate Poisson regression analysis was performed to evaluate the association between covariates and ICD therapies. RESULTS: and 69% had an undetectable viral load. The median follow-up was 6.4 years. HIV-infected subjects had both a higher number of appropriate ICD shocks or antitachycardia pacing (ATP) therapy per person-year as well as a higher number of inappropriate ICD shocks per person-year than uninfected controls (1.512 vs. 0.590 and 0.122 vs. 0.0166, respectively, p < .001 for both comparisons). After multivariate adjustment, the presence of detectable/unsuppressed viral load at the time of ICD implantation was an independent predictor of both of the following in HIV-infected subjects: (1) appropriate ICD discharge (p = .004), and (2) appropriate ICD discharge or appropriate ATP therapy (p < .001). CONCLUSION: HIV-infected subjects had a higher number of appropriate ICD discharge or ATP therapy per person-year than matched uninfected controls.
Narla et al. (2022) conducted a cohort in HIV infection with ICD implantation (n=71). HIV infection vs. Uninfected controls was evaluated on Appropriate ICD shocks or antitachycardia pacing (ATP) therapy per person-year (p=<.001). HIV-infected patients had a higher rate of appropriate ICD shocks or antitachycardia pacing therapy per person-year than matched uninfected controls.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: