Key result
Epicardial ICDs in infants and toddlers yield 100% survival, with ~20% receiving appropriate shocks.
Why the study?
Limited data exist regarding implantable cardioverter defibrillator usage in infants and toddlers.
Cohort (n=15)
No
Epicardial ICDs can be safely and effectively placed in infants and toddlers ≤3 years old for sudden death prevention, yielding good midterm outcomes with appropriate shocks delivered in 20% of patients and no inappropriate shocks.
Answers are grounded in this paper and cite their sources.
Adds limited descriptive data on ICD use in infants and toddlers; leaves open optimal strategies pending prospective studies.
BACKGROUND: Limited data exist regarding implantable cardioverter defibrillator (ICD) usage in infants and toddlers. This study evaluates ICD placement indications, procedural techniques, programming strategies, and outcomes of ICDs in infants and toddlers. METHODS: This is a single-center retrospective review of all patients ≤3 years old who received an ICD from 2009 to 2021. RESULTS: Fifteen patients received an ICD at an age of 1.2 years (interquartile range [IQR], 0.1-2.4; 12 [80%] women; weight, 8.2 kg [IQR, 4.2-12.6]) and were followed for a median of 4.28 years (IQR, 1.40-5.53) or 64.2 patient-years. ICDs were placed for secondary prevention in 12 patients (80%). Diagnoses included 8 long-QT syndromes (53%), 4 idiopathic ventricular tachycardias/ventricular fibrillations (VFs; 27%), 1 recurrent ventricular tachycardia with cardiomyopathy (7%), 1 VF with left ventricular noncompaction (7%), and 1 catecholaminergic polymorphic ventricular tachycardia (7%). All implants were epicardial, with a coil in the pericardial space. Intraoperative defibrillation safety testing was attempted in 11 patients (73%), with VF induced in 8 (53%). Successful restoration of sinus rhythm was achieved in all tested patients with a median of 9 (IQR, 7.3-11.3) J or 0.90 (IQR, 0.68-1.04) J/kg. Complications consisted of 1 postoperative chylothorax and 3 episodes of feeding intolerance. VF detection was programmed to 250 (IQR, 240-250) ms with first shock delivering 10 (IQR, 5-15) J or 1.1 (IQR, 0.8-1.4) J/kg. Three patients (20%) received appropriate shocks for ventricular tachycardia/VF. No patient received an inappropriate shock. There were 2 (13%) ventricular lead fractures (at 2.6 and 4.2 years post-implant), 1 (7%) pocket-site infection, and 2 (13%) generator exchanges. All patients were alive, and 1 patient (7%) received a heart transplant. CONCLUSIONS: ICDs can be safely and effectively placed for sudden death prevention in infants and toddlers with good midterm outcomes.
No takes yet. Share an insight, caveat, or question.
A 2022 study conducted a cohort in Ventricular arrhythmias or high risk of sudden cardiac death (n=15). Implantable cardioverter defibrillator (ICD) was evaluated on Appropriate shocks for ventricular tachycardia/VF. Epicardial implantable cardioverter defibrillators in infants and toddlers provided effective sudden death prevention, with 20% receiving appropriate shocks and 100% survival over 4.28 years.
Synapse has enriched 10 related papers. Consider them for comparative context: