Key result
EV-ICD implantation achieves 100% acute success without surgical complications in pediatric patients.
Why the study?
The extravascular implantable cardioverter-defibrillator (EV-ICD) is an appealing option for pediatric patients because it avoids intravascular hardware and offers a smaller generator and anti-tachycardia pacing, but reports of implant technique and performance in children and adolescents were limited.
What are the acute surgical outcomes, electrical performance, and short-term complications of extravascular implantable cardioverter-defibrillator (EV-ICD) implantation in pediatric and adolescent patients?
Observational (n=12)
No
What are the acute surgical outcomes, electrical performance, and short-term complications of extravascular implantable cardioverter-defibrillator (EV-ICD) implantation in pediatric and adolescent patients?
EV-ICD implantation in pediatric and adolescent patients is feasible and acutely safe when incorporating anatomical surgical adaptations and routine defibrillation threshold testing.
Answers are grounded in this paper and cite their sources.
EV-ICD may be feasible in select pediatric patients with adaptations; leaves open long-term efficacy and safety in larger cohorts.
Background The extra-vascular implantable cardioverter defibrillator (EV-ICD) (Medtronic Inc, Minneapolis, MN, USA) is safe and effective among adult patients. By avoiding intravascular hardware and offering a smaller generator and anti-tachycardia pacing, it is an appealing option for pediatric patients but reports of implant technique and performance in children are limited. Objective To report patient characteristics, outcomes, and complications of EV-ICD implant in children and adolescents, to describe adaptations to manufacturer recommended implant techniques, and report available post-implant data. Methods Single center retrospective review of pediatric patients who underwent EV-ICD implantation at Children's Hospital Colorado from January 2024 through November 2025. Patient characteristics, acute surgical outcomes and complications, electrical performance, and follow-up data were summarized. Results 12/12 patients (3/12 female (25%, median age 16 years [12-18], weight 59 kg [48-87])) had acutely successful implants without surgical complications. 9/12 (75%) had difficult to palpate xiphoid processes, but an oblique incision along the left costal margin allowed for uncomplicated retrosternal access. Procedural characteristics included median procedure time of 146 min (110-156), fluoroscopy time of 5.8 min (3.8-6.3), mean R-wave sensing of 6.6 mV (±3.2 mV), with 0/12 patients demonstrating P-wave oversensing. 11/12 underwent DFT testing, with 2 initial DFT failures – one for undersensing of VF that required a change in sensing vector and one which required repositioning of the generator to improve shock vector. Median follow-up of 8 months (4-15) demonstrated 3 appropriate shocks in two patients (for VF and monomorphic VT) and 2 complications (16.7%): post-operative pain requiring extended observation in one patient and an inappropriate shock for sinus tachycardia in another. Conclusions EV-ICD implantation among children and adolescents was acutely successful without procedural complications. Pediatric specific adaptations for anatomic variants (non-palpable xiphoid process, sternal ossification centers, sternal curvature) may aid in implant procedure. DFT testing is recommended to ensure appropriate sensing and shock vectors. Post-implant device performance was stable and multiple appropriate shocks were successful.
No takes yet. Share an insight, caveat, or question.
Nash et al. (2026) conducted an observational in Pediatric and adolescent patients requiring an implantable cardioverter-defibrillator (n=12). Extravascular implantable cardioverter-defibrillator (EV-ICD) implantation was evaluated on Acute implant success without surgical complications. EV-ICD implantation in 12 pediatric patients achieved 100% acute success without surgical complications, with 3 appropriate shocks and a 16.7% complication rate over median 8-month follow-up.