Key result
Extravascular ICD implantation achieves 100% success in children with low early complication rates.
Why the study?
The extravascular implantable cardioverter-defibrillator (EV-ICD) is a novel substernal ICD with limited reports in children and adolescents.
Does extravascular implantable cardioverter-defibrillator (EV-ICD) implantation provide reliable implantation and short-term device performance in children and adolescents <18 years?
Observational (n=15)
No
Does extravascular implantable cardioverter-defibrillator (EV-ICD) implantation provide reliable implantation and short-term device performance in children and adolescents <18 years?
Answers are grounded in this paper and cite their sources.
This largest international single-centre paediatric series demonstrates that extravascular ICDs can be reliably implanted in children as young as 4 years old (17.3 kg) with stable short-term sensing and appropriate shock delivery.
The extravascular implantable cardioverter-defibrillator (EV-ICD) is a novel substernal ICD with limited reports in children and adolescents. Single-centre review of EV-ICDs in children <18-years. Clinical, peri-procedural, and electrocardiographic data reviewed, including device performance in routine exercise stress testing. Median follow-up 10.9 (IQR 7.5 – 12.5) months. Between April 2025-April 2026, 15 EV-ICDs were successfully implanted by a single cardiothoracic surgeon: median age 13.8-years, IQR 10.2–16.4; 11 (73%) male; weight 52 kg, IQR 38.2–64.1; BSA 1.6 m 2 , IQR 1.2–1.8. Smallest child was 4-years, 17.3 kg (BSA 0.7 m 2 ). Indications were primary ( n = 10) and secondary ( n = 5) prevention. Underlying diagnoses included cardiomyopathy ( n = 9), channelopathy ( n = 3). One child had a previous sternotomy and another severe chest-wall deformity. Median anaesthetic time was 148 min (IQR 130–161), surgical time 68 min (59–88), fluoroscopy time 2.6 min (2.2–3.8). Median 3 (IQR 2–4) day hospitalisation. Two children experienced early complications. In one child, defibrillator threshold testing terminated ventricular fibrillation (VF) into ventricular tachycardia (VT) requiring CPR and a decision to use external cardioversion. The child fully recovered. One adolescent experienced profound bradycardia and syncope early post anaesthesia. R-wave amplitude at implant (5mV, IQR 4.2 – 6.7), discharge (5, IQR 2.8 – 8.1) and latest follow-up (5.8, IQR 3.2 – 13.7) was stable. Device sensing during routine exercise stress testing has been appropriate. There has been one successful appropriate high voltage therapy and no inappropriate high voltage therapy. This is the largest single-centre series internationally, demonstrating reliable paediatric EV-ICD implantation and short-term device performance. Ongoing paediatric surveillance is necessary.
No takes yet. Share an insight, caveat, or question.
Luxford et al. (2026) conducted an observational in Paediatric primary or secondary prevention of sudden cardiac death (cardiomyopathy, channelopathy) (n=15). Extravascular implantable cardioverter-defibrillator (EV-ICD) was evaluated on Successful implantation, peri-procedural complications, and short-term device performance. Extravascular ICD implantation in 15 children achieved 100% implant success with 2 early complications, 1 appropriate therapy, and 0 inappropriate shocks over 10.9 months.