Key result
Model LA-201 lead adapter is linked to ~5-fold higher failure risk versus other adapters.
Why the study?
The lack of universal industry standards and variety of ICD manufacturers necessitate the use of lead adapters, whose performance and safety require evaluation.
Does the model of ICD pacing/sensing lead adapter affect the rate of device failure?
Cohort (n=81)
Does the model of ICD pacing/sensing lead adapter affect the rate of device failure?
ICD pacing/sensing lead adapters are a potential source of complications, with model LA-201 showing a significantly higher long-term failure rate compared to other models.
Answers are grounded in this paper and cite their sources.
May warrant avoiding LA-201 adapters pending validation; leaves open prospective assessment of lead durability.
The wide variety of implantable defibrillators (ICDs) available from different manufactures and the lack of universal industry standards has resulted in the frequent need for lead adapters at time of ICD implant or change. We analyzed the performance of 81 consecutive ICD sensing/pacing lead adapters used between 1988 and 1993. A total of 66 adapters was used for new epicardial systems, and 15 adapters served as lead connectors during ICD generator replacement. Pacing/sensing lead adapters used were: model LA-201 (n = 28; 34.5%); model 030-308 (n = 26; 32%); model 5866-24 (n = 15; 18.5%); and miscellaneous (n = 12; 15%). After a mean follow-up of 21 +/- 16 months, nine pacing/sensing lead adapters had documented or strongly suspected failure. Most often pacing/sensing lead adapters presented clinically as frequent aborted shocks. Actuarial probability of freedom from failure for model LA-201 was 83% at 1 year, and 72% at 2 and 3 years; this was poorer than for the other sensing leads combined (P = 0.01; hazard ratio = 4.92; 95% confidence intervals = 1.2-20; log-rank test). In conclusion, pacing/sensing lead adapters are a potential source of ICD system complications. Performance is dissimilar among different models; specifically, model LA-201 may not be safe in the long-term, and patients with this lead adapter need to be closely monitored.
No takes yet. Share an insight, caveat, or question.
Sgarbossa et al. (1996) conducted a cohort in Patients requiring implantable defibrillator (ICD) pacing/sensing lead adapters (n=81). Model LA-201 pacing/sensing lead adapter vs. Other sensing leads combined was evaluated on Adapter failure (HR 4.92, 95% CI 1.2-20, p=0.01). The model LA-201 pacing/sensing lead adapter was associated with a higher risk of failure compared to other sensing leads combined.
Synapse has enriched 10 related papers. Consider them for comparative context: