Key result
Antitachycardia pacing successfully terminates ~91% of spontaneous VT episodes.
Why the study?
The efficacy of antitachycardia pacing by a third generation implantable cardioverter defibrillator in converting induced versus spontaneous ventricular tachycardia required evaluation.
Does antitachycardia pacing by the VENTAK PRx pulse generator terminate episodes of ventricular tachycardia in patients with a history of VT or sudden cardiac death?
Observational (n=70)
Does antitachycardia pacing by the VENTAK PRx pulse generator terminate episodes of ventricular tachycardia in patients with a history of VT or sudden cardiac death?
Antitachycardia pacing by the VENTAK PRx device successfully terminates the vast majority of both induced and spontaneous ventricular tachycardia episodes.
Initial US experience supports multiprogrammable ICD feasibility in CAD-dominant patients; leaves open optimal programming and long-term efficacy.
Seventy consecutive patients received the first VENTAK PRx pulse generators (Cardiac Pacemakers, Inc.) implanted in the United States. This multiprogrammable device has therapeutic options that include: (1) antitachycardia pacing; (2) low energy cardioversion; (3) defibrillation shocks; and (4) bradycardia pacing. There were 60 males and 10 females with a mean age of 65.3 +/- 9.4 years. The anatomical diagnoses were coronary artery disease in 55 patients, cardiomyopathy in 7 patients, congenital heart disease in 2 patients, and miscellaneous disease in the remaining 6 patients. Thirty-six percent had a history of sudden cardiac death and 90% had documented monomorphic ventricular tachycardia. The mean ejection fraction was 32.7% +/- 12.2%. Thirty-three (49.3%) had an ejection fraction < or = 30%. Electrophysiological testing was done preimplant, predischarge, and 1 to 2 months postimplant to define a specific electrical therapy and evaluate the efficacy of the device. Two hundred ninety-three of 367 (80%) episodes of induced ventricular tachycardia were successfully terminated by the VENTAK PRx programmed for antitachycardia pacing. There were 1,794 total therapy episodes for spontaneous ventricular tachycardia; 91% (1,641 episodes) were terminated by antitachycardia pacing and 153 episodes were converted by shocks during a minimal 6-month follow-up per patient. We conclude that documentation of a successful antitachycardia pacing modality in the electrophysiology laboratory predicts conversion of spontaneous episodes of ventricular tachycardia. Furthermore, antitachycardia pacing by the VENTAK PRx can terminate the majority of episodes of ventricular tachycardia.
No takes yet. Share an insight, caveat, or question.
Porterfield et al. (1993) conducted an observational in Ventricular tachycardia (n=70). VENTAK PRx pulse generator was evaluated on Termination of spontaneous ventricular tachycardia episodes by antitachycardia pacing. Antitachycardia pacing by the VENTAK PRx pulse generator successfully terminated 80% of induced and 91% of spontaneous ventricular tachycardia episodes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: