Key result
Moderate-to-severe LVOT calcium is linked to greater THV inflow eccentricity, but circularity is preserved at leaflets.
Why the study?
Left ventricular outflow tract calcium location and degree may affect transcatheter heart valve deployment and impact flow patterns and valve durability in TAVR patients.
Does moderate/severe LVOT calcium impact valve geometry and hemodynamics in patients undergoing TAVR with self-expanding valves?
Cohort (n=107)
Yes
Does moderate/severe LVOT calcium impact valve geometry and hemodynamics in patients undergoing TAVR with self-expanding valves?
Self-expanding transcatheter heart valves maintain proper leaflet circularity and hemodynamic function despite inflow deformities caused by moderate/severe LVOT calcium.
Answers are grounded in this paper and cite their sources.
Should not yet change TAVR planning for LVOT calcium; leaves open long-term durability implications of inflow eccentricity.
OBJECTIVES: Evaluate transcatheter heart valve (THV) geometry according to left ventricular outflow tract (LVOT) calcium degree and its impact on hemodynamics and outcomes in patients undergoing transcatheter aortic valve replacement (TAVR) with a contemporary self-expanding THV. BACKGROUND: LVOT calcium remains challenging for contemporary THVs. LVOT calcium location and degree may affect THV deployment and impact flow patterns and shear stress, accelerating THV degeneration. METHODS: EPROMPT (CoreValve Evolut Pro Prospective Registry; NCT03423459) is a prospective, investigator-initiated, multicenter registry of patients undergoing TAVR using CoreValve Evolut PRO/PRO + THVs. A total of 107 patients were enrolled in EPROMPT's computed tomography (CT) cohort between January 2018 and October 2021. These patients underwent follow-up CT scan 30 days post-TAVR. We analyzed THV geometry and its interaction with the aortic root following deployment using 30-day post-TAVR CT in patients with none/mild versus moderate/severe LVOT calcium. RESULTS: Thirty-day THV inflows were less eccentric in the short axis in patients with none/mild versus moderate/severe LVOT calcium (1.16 ± 0.09 vs. 1.21 ± 0.12; p = 0.007). THV became more circular and was similar between both cohorts at the THV waist (1.08 ± 0.06 vs. 1.09 ± 0.11; p = 0.551), leaflet tips (1.03 ± 0.04 vs. 1.05 ± 0.09; p = 0.299), and THV outflow (1.04 ± 2.2 vs. 1.03 ± 2.7; p = 0.143). Thirty-day > mild paravalvular leak was low in both cohorts (1.5% vs. 2.4%; p = 0.724); mean gradients were similar (7.7 ± 3.6 vs. 7.7 ± 3.4 mmHg; p = 0.955). CONCLUSIONS: Despite inflow deformities observed in patients with moderate/severe LVOT calcium, Evolut PRO/PRO + conforms to elliptical aortic annuli, maintaining circularity and proper function at the leaflets and outflow, even in patients with moderate/severe LVOT calcium.
No takes yet. Share an insight, caveat, or question.
Medranda et al. (2022) conducted a cohort in Aortic stenosis undergoing TAVR (n=107). Moderate/severe LVOT calcium vs. None/mild LVOT calcium was evaluated on 30-day THV inflow eccentricity in the short axis (p=0.007). Moderate/severe LVOT calcium was associated with more eccentric 30-day THV inflows compared to none/mild calcium (1.21 vs 1.16; p=0.007), though circularity was maintained at the leaflets.
Synapse has enriched 10 related papers. Consider them for comparative context: