Answers are grounded in this paper and cite their sources.
High-frequency electric welding (HFEW) technology holds great promise for small intestine end-to-end anastomoses, but existing energy generators lack feedback mechanisms to quantify energy transfer and welding strength. This study proposes the dynamic impedance ratio ( η = R f / R min ) as a novel feedback parameter to standardize intraoperative assessment while mitigating individual variability. Systematic investigations were conducted through ex vivo porcine small intestine end-to-end welding experiments, analyzing the effects of compression pressure (50 to 400 kPa), voltage (40 to 60 V), and welding duration (10 to 20 s) on dynamic impedance. A short pulse signal measures how mechanical pressure affects tissue impedance. Under 150 to 350 kPa, impedance drops to 40% to 70% of its initial value. Results demonstrated that compression pressure reduced initial impedance by decreasing tissue thickness, while voltage and duration modulated dynamic impedance through thermal denaturation and dehydration. η shows a strong positive correlation with burst pressure, effectively predicting weld strength. A higher η indicates greater burst pressure, and when η ≥ 2, the burst pressure test achieves a 100% success rate. As a predictor of welding strength, dynamic impedance ratio can predict weld strength without destroying the anastomosis tissue, which has wide application in the future. It will further improve the accuracy of energy regulation and further promote the use of HFEW in clinical applications.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: