Key result
Novel snare-in-delivery-system technique enables successful retrieval of depleted Micra TPS during replacement.
Why the study?
A new technique for elective replacement of a leadless pacemaker due to battery depletion was needed to avoid device abandonment and related complications.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
A novel snare loop technique allows for the successful retrieval of a depleted leadless pacemaker during replacement, avoiding device abandonment in the right ventricle.
We report the first case of new technique of replacement of a Micra TPS, due to battery depletion. A 38-year-old patient was admitted due to battery depletion of a TPS, after 44 months of regular pacemaker functioning. After routine implantation of a new TPS, we use a snare loop inserted in the delivery system to capture the old TPS. We believe this approach a good option not to abandon the depleted device, to avoid possible electrical interference or space occupation in right ventricle. This new approach allows to change the strategy during procedure and does not increase significantly the procedure costs.
No takes yet. Share an insight, caveat, or question.
Filippo et al. (2021) conducted a case report in Battery depletion of a Micra TPS (leadless pacemaker) (n=1). Replacement of Micra TPS using a snare loop to capture the old device was evaluated on Successful replacement and retrieval of old TPS. A new technique using a snare loop inserted in the delivery system successfully captured and removed a depleted Micra TPS during the implantation of a new device.
Synapse has enriched 10 related papers. Consider them for comparative context: