Key result
A rapidly developed 15-channel flexible foot and ankle MRI coil demonstrated comparable image quality to a commercially available 8-channel RF coil in a healthy volunteer.
Highly decoupled RF coil elements enable rapid prototyping of flexible, application-specific MRI receive coils with comparable image quality to commercial rigid coils.
This is the authors' abstract. We don't add key points for this paper.
Over the last 30 years, there have been dramatic changes in phased array coil technology leading to increasing channel density and parallel imaging functionality. Current receiver array coils are rigid and often mismatched to patient's size. Recently there has been a move towards flexible coil technology, which is more conformal to the human anatomy. Despite the advances of so-called flexible surface coil arrays, these coils are still relatively rigid and limited in terms of design conformability, compromising signal-to-noise ratio (SNR) for flexibility, and are not designed for optimum parallel imaging performance. The purpose of this study is to report on the development and characterization of a 15-channel flexible foot and ankle coil, rapidly designed and constructed using highly decoupled radio-frequency (RF) coil elements. Coil performance was evaluated by performing SNR and g-factor measurements. In vivo testing was performed in a healthy volunteer using both the 15-channel coil and a commercially available 8-channel foot coil. The highly decoupled elements used in this design allow for extremely rapid development and prototyping of application-specific coils for different patient sizes (adult vs child) with minimal additional design consideration in terms of coil overlap and geometry. Image quality was comparable to a commercially available RF coil.
Answers are grounded in this paper and cite their sources.
Supports rapid prototyping of flexible MRI coils; hypothesis-generating from single-volunteer data and requires clinical validation.
No takes yet. Share an insight, caveat, or question.
Collick et al. (2020) studied Healthy volunteer (n=1). 15-channel flexible foot and ankle coil vs. Commercially available 8-channel foot coil was evaluated on Signal-to-noise ratio (SNR), g-factor measurements, and image quality. A rapidly developed 15-channel flexible foot and ankle MRI coil demonstrated comparable image quality to a commercially available 8-channel RF coil in a healthy volunteer.