Answers are grounded in this paper and cite their sources.
Contamination of X-ray mirror surfaces by carbon is a common issue that can significantly degrade the optical performance of the instrument. The effects can be severe at photon energies near the carbon K-edge (ca. 300 eV), where the X-rays are strongly attenuated, but also significant at higher photon energies where the carbon coating affects the reflectivity and surface shape of the mirrors. [1] The Swiss Light Source has typically relied on in-situ plasma cleaning to control mirror contamination and the PolLux scanning transmission X-ray microscopy (STXM) beamline has also been employing further contamination reduction strategies in recent years. In particular, in 2014 we installed a 1 × 10 −8 mbar background pressure of O 2 on the PolLux first mirror chamber. We present a history of efforts to control optical contamination at the PolLux beamline and report on the observed efficiencies of the different processes employed both for the in-vacuum optics and critical components of the frequently vented STXM experiment chamber.
No takes yet. Share an insight, caveat, or question.
Watts et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: