Answers are grounded in this paper and cite their sources.
We are developing kilo-pixel arrays of transition-edge sensor (TES) microcalorimeters for use in future laboratory and space based X-ray astrophysics experiments. These arrays are required to achieve an energy resolution of ΔEFWHMCand magnetic field uniformity can be reduced whilst maintaining the target energy resolution. Results show ΔEFWHM≈ 2.0 - 2.5 eV FWHM at a photon energy of 6 keV, measured across several pixels. Despite larger than typical TCvariation across the detector, time division multiplexed readout of 30 common electrically biased pixels in the 32 × 32 array (in a 2 column × 16 row configuration) show an average ΔEFWHM= 3.0 ± 0.3 eV.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2014) studied this question.