An enantiomeric separation of meclizine enantiomers by liquid chromatography with tandem mass spectrometry LC-MS method was developed and validated for the analysis of Meclizine enantiomers. Enantiomeric resolution of the drug products were successfully achieved on a Phenomenex® lux cellulose 1 C18 (250 mm × 4.6 mm i.d, 5 µm particle size) column with mobile phase consisting of acetonitrile: 5 mM ammonium format pH (5.5) adjusted with formic acid (90:10) (v/v), and a flow rate of 0.4 mL/min. The developed method provided linear responses within the concentration range 1–5 ng/mL, and regression analysis showed a correlation coefficient value (r2) of 0.999. The optimized mobile phase separated (+) Meclizine at 1.58 min and (-) Meclizine at 2.20 min, respectively. The LC/MS method was validated as per ICH guidelines with respect to specificity, precision, linearity and robustness. Limit of detection (LOD) and limit of quantification (LOQ) were found to be 1.0 ng/mL and 5.0 ng/mL respectively. The proposed method is suitable for analysis of meclizine enantiomers in pharmaceutical formulations and quality control analysis.