Key result
Pulsatile flow reduces stent displacement force by ~12% compared to static conditions in bench testing.
Why the study?
The effect of vessel material properties and pulsatile wall motion on stent fixation after endovascular abdominal aortic aneurysm repair had not been previously investigated.
Does pulsatile flow and vessel compliance reduce the displacement force of proximal stents in endovascular grafts?
Does pulsatile flow and vessel compliance reduce the displacement force of proximal stents in endovascular grafts?
In an in vitro model, more compliant vessel walls and pulsatile wall motions significantly reduced the fixation force of passively fixated aortic stents, highlighting potential mechanisms for endograft migration.
Answers are grounded in this paper and cite their sources.
May reduce proximal stent fixation in compliant vessels with pulsatile flow; leaves open mechanisms of clinical endograft migration.
Migration is a serious failure mechanism associated with endovascular abdominal aortic aneurysm (AAA) repair (EVAR). The effect of vessel material properties and pulsatile wall motion on stent fixation has not been previously investigated. A proximal stent from a commercially available stent graft was implanted into the proximal neck of silicone rubber abdominal aortic aneurysm models of varying proximal neck stiffness (β=25.39 and 20.44). The stent was then dislodged by placing distal force on the stent struts. The peak force to completely dislodge the stent was measured using a loadcell. Dislodgment was performed at ambient pressure with no flow (NF) and during pulsatile flow (PF) at pressures of 120/80 mmHg and 140/100 mmHg to determine if pulsatile wall motions affected the dislodgement force. An imaging analysis was performed at ambient pressure and at pressures of 120 mmHg and 140 mmHg to investigate diameter changes on the model due to the radial force of the stent and internal pressurisation. Stent displacement forces were ~50% higher in the stiffer model (7.16-8.4 N) than in the more compliant model (3.67-4.21 N). The mean displacement force was significantly reduced by 10.95-12.83% from the case of NF to the case of PF at 120/80 mmHg. A further increase in pressure to 140/120 mmHg had no significant effect on the displacement force. The imaging analysis showed that the diameter in the region of the stent was 0.37 mm greater in the less stiff model at all the pressures which could reduce the fixation of the stent. The results suggest that the fixation of passively fixated aortic stents could be comprised in more compliant walls and that pulsatile motions of the wall can reduce the maximum stent fixation.
No takes yet. Share an insight, caveat, or question.
Corbett et al. (2010) studied Abdominal aortic aneurysm. Pulsatile flow and varying proximal neck stiffness vs. No flow and stiffer model was evaluated on Peak force to completely dislodge the stent. Stent displacement forces were ~50% higher in stiffer models, and pulsatile flow significantly reduced mean displacement force by 11-13% compared to no flow.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: