Key result
Magna Ease aortic valve replacement shows ~81% freedom from all-cause mortality at 8 years.
Why the study?
Long-term outcomes of the Carpentier-Edwards PERIMOUNT Magna Ease valve after aortic valve replacement were not previously reported.
What are the long-term safety, hemodynamic, and functional outcomes of aortic valve replacement with the Carpentier-Edwards PERIMOUNT Magna Ease valve in adults?
Cohort (n=258)
Yes
What are the long-term safety, hemodynamic, and functional outcomes of aortic valve replacement with the Carpentier-Edwards PERIMOUNT Magna Ease valve in adults?
The Carpentier-Edwards PERIMOUNT Magna Ease valve demonstrates satisfactory safety, durability, and sustained hemodynamic and functional improvements at 8 years post-AVR.
8-year outcomes support satisfactory safety and hemodynamics with Magna Ease AVR; extends single-arm data but leaves open comparative effectiveness.
Introduction The Carpentier-Edwards PERIMOUNT Magna Ease valve is a third-generation bioprosthesis for aortic valve replacement (AVR). This is a postapproval study reporting on its 8-year outcomes. Methods Adults undergoing AVR with the Magna Ease valve between October 2007 and December 2012 were enrolled for this prospective, nonrandomized, single-arm, and multicenter study. Assessments occurred preoperatively, at hospital discharge, 6 months, 1 year, and annually thereafter for up to 8 years. Outcomes included safety endpoints, hemodynamic performance, and New York Heart Association (NYHA) functional class. Results Of the 258 study patients, 67.5% were in NYHA Class I or II, and 32.5% were in NYHA Class III or IV at baseline. Concomitant procedures were performed in 44.2%. Total follow-up was 1597.6 patient-years, and median follow-up was 7 years (interquartile range: 5.5–8.0 years). Eight years following AVR, the functional class remained improved from baseline with 93.9% in NYHA Class I/II and 6.1% in NYHA Class III; 38 deaths had occurred, 8 of which were valve related; freedom from all-cause mortality was 80.7% (95% confidence intervals: 74.9, 86.4); freedom from valve-related mortality was 95.8% (92.8, 98.8); freedom from reintervention, explant, major bleeding events, and structural valve deterioration was 89.8% (85.1, 94.6), 94.8% (91.7, 97.9), 85.1% (80.0, 90.1), and 90.1% (84.7, 95.4), respectively; effective orifice area was 1.5 ± 0.5 cm2, the mean gradient was 14.8 ± 8.3 mmHg, and 88.6% of patients had no or trivial aortic regurgitation. Conclusions This study demonstrated satisfactory safety and sustained hemodynamic and functional improvements at 8 years following AVR with the Magna Ease valve.
No takes yet. Share an insight, caveat, or question.
Tsui et al. (2022) conducted a cohort in Aortic valve replacement (n=258). Carpentier-Edwards PERIMOUNT Magna Ease valve was evaluated on Freedom from all-cause mortality (95% CI 74.9-86.4). Aortic valve replacement with the Carpentier-Edwards PERIMOUNT Magna Ease valve demonstrated 80.7% freedom from all-cause mortality (95% CI 74.9-86.4) and sustained functional improvements at 8 years.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: