Why the study?
The study aimed to compare the long-term clinical and hemodynamic outcomes of aortic valve replacement using Carpentier-Edwards Perimount versus Perimount Magna valves.
Does aortic valve replacement with Perimount Magna valves improve overall survival compared to Perimount valves in patients undergoing aortic valve replacement?
Does aortic valve replacement with Perimount Magna valves improve overall survival compared to Perimount valves in patients undergoing aortic valve replacement?
Although the Perimount Magna valve led to decreased aortic valve pressure gradients, the older Perimount valve was associated with significantly better overall and event-free survival.
May support Magna over Perimount selection; leaves open need for randomized confirmation in AVR.
BACKGROUND: The present study aimed to compare the long-term clinical and hemodynamic outcomes of aortic valve replacement using Carpentier-Edwards Perimount (Perimount) or Perimount Magna (Magna) valves. METHODS: We enrolled 430 patients who underwent aortic valve replacements with Perimount (n=58) or Magna (n=372) valves [1998-2013]. Multivariable and inverse probability of treatment weight (IPTW) analyses were performed. RESULTS: Magna 76%±4%; P=0.02; hazard ratio (HR): 0.37 for the Perimount group; 95% confidence interval (CI): 0.17-0.83]. Multivariable analysis of the overall survival identified Perimount valve use as a protective factor (P=0.009; HR: 0.32; 95% CI: 0.14-0.75). Independent risk factors of overall survival were older age, male sex, higher preoperative left ventricular mass index, lower ejection fraction, lower aortic valve pressure gradient, and lower haemoglobin. After applying IPTW, overall survival was again found to be significantly longer in the Perimount group (P=0.04; HR: 0.43; 95% CI: 0.20-0.93). Event-free survival was also better in the Perimount group (P=0.006; HR: 0.38; 95% CI: 0.19-0.75). However, the Magna group had significantly lower aortic valve pressure gradients at one year and five years postoperative. CONCLUSIONS: Although Magna use led to decreased aortic valve pressure gradients at follow-up, overall and event-free survival rates were significantly better with use of the Perimount valve. Additional and larger studies are needed to confirm these results.
No takes yet. Share an insight, caveat, or question.
Chang et al. (2019) studied this question.