Answers are grounded in this paper and cite their sources.
BACKGROUND: During continuous renal replacement therapy (CRRT) with regional citrate anticoagulation (RCA), blood flow (Qb) might affect vascular access dysfunction (AD) and, thereby, circuit life. The three sites shared similar RCA protocols but differed in Qb (120-130 vs 150-200 mL/h). Non-RCA circuits were compared with RCA circuits in which the impact of Qb was also assessed. RESULTS: About 3,981,906 min of circuit pressures were analysed in 2568 circuits in 567 patients. High-Qb RCA was associated with more extreme pressures, and greater AD (IRR 3.7 (1.93-7.08) as well as reduced filter life 21.1 (10.2-42.6) vs 27.0 (14.8-41.6) h). AD in high-Qb RCA circuits was associated with a 49% reduction in filter life, versus 24% reduction in low-Qb RCA, associated with a rise in the rate of increase in transfilter pressure. CONCLUSIONS: High-Qb RCA-CRRT was associated with greater access dysfunction, earlier filter loss and increased haemodynamic impacts of access dysfunction, suggesting low-Qb RCA-CRRT may improve circuit mechanics, function and longevity.
No takes yet. Share an insight, caveat, or question.
Sansom et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: