Key result
Resistance-guided ICD shock pulse duration and tilt programming optimizes theoretical defibrillation efficacy.
Why the study?
Manufacturers of ICDs do not commonly advise how to program shock pulse duration and tilt for optimal clinical benefit.
How should pulse duration and tilt be programmed in ICDs to optimize defibrillation based on lead system resistance?
How should pulse duration and tilt be programmed in ICDs to optimize defibrillation based on lead system resistance?
Programming ICD shock pulse duration and tilt based on lead system resistance may theoretically optimize defibrillation efficacy.
Answers are grounded in this paper and cite their sources.
Theoretical ICD waveform optimization by lead resistance should not yet guide practice; leaves open need for clinical validation trials.
Implantable cardioverter defibrillators (ICDs) are available with independently programmable duration and tilt of the shock pulse waveform. Manufacturers do not, however, commonly advise how these parameters can be programmed for optimal clinical benefit. From theoretical considerations, the author recommends programming both parameters based on the measured lead system resistance R into which the shock is delivered. Assuming that the defibrillation pulse decline below the defibrillation threshold rheobase is undesirable because of the possibility of refibrillation. Mathematical relationships expressing optimal pulse duration and tilt as functions of the output time constant can be derived that are valid for monophasic pulses and the first phase of biphasic pulses. Two ICD manufacturers provide for programmable tilt (Medtronic GEM III, atrial channel) or both tilt T and pulse duration PD. (St. Jude Medical newest devices). Considering its output capacitance, it is recommended that the Medtronic Gem III should be programmed for T = 50% when R < 75 omega and 65% when R < 38 omega. The author considers programming tilt to 30% or 40% useless in clinical conditions. By the same reasoning, he recommends that the newer St. Jude Medical ICDs should be programmed to T = 50% if R < 75 omega and 60% if R < 41 omega, and PD = 5.5, 5.0, 4.5, 4.0, 3.5, and 3.0 ms for R < 75, 73, 62, 51, 40, and 32 omega, respectively. PD = 6 ms was considered clinically unrealistic. Programmable shock pulse duration and tilt are useful in optimizing defibrillation, but it is suggested that this can best be accomplished by programming these parameters with the guidance of theory as described in this discussion.
No takes yet. Share an insight, caveat, or question.
Werner E. Irnich (2003) studied Implantable cardioverter defibrillators. Programming pulse duration and tilt based on lead system resistance was evaluated. Programming ICD shock pulse duration and tilt based on measured lead system resistance is theoretically recommended to optimize defibrillation.
Synapse has enriched 10 related papers. Consider them for comparative context: