Key result
Cardiac 3T MRI proves safe and feasible in a patient with a leadless pacemaker.
Why the study?
Is 3.0 Tesla cardiac MRI safe and diagnostically useful in a patient with a leadless cardiac pacemaker?
Case Report (n=1)
No
Is 3.0 Tesla cardiac MRI safe and diagnostically useful in a patient with a leadless cardiac pacemaker?
This first report demonstrates that 3.0 Tesla cardiac MRI can be performed safely in a patient with a leadless cardiac pacemaker without significant loss of diagnostic quality or device malfunction.
Answers are grounded in this paper and cite their sources.
Supports 3T MRI feasibility with leadless pacemakers in select cases; leaves open need for systematic safety data.
This is to the best of our knowledge the first report of cardiac magnetic resonance imaging (MRI) in a patient with a leadless cardiac pacemaker (LCP). Imaging was performed to rule out myocarditis in a 77-year-old male patient who had undergone LCP implantation (MicraTM, Medtronic) for atrial fibrillation with bradycardia 20 months before (Panel A). After a precise check of the functional parameters, the device was programmed to the MRI mode (V00 with a fixed rate of 80 b.p.m.). The MRI was performed in a long bore 3.0 Tesla magnet (Magnetom®Skyra, Siemens, Erlangen, Germany) with a maximal specific absorption rate of 1.5 W/kg. During MRI, the patient was monitored continuously by electrocardiogram and pulse oximetry. The cardiac MRI showed metallic artefacts at the apex of the heart due to the implanted LCP in the apex of the right ventricle and at the sternum due to sternal wires after cardiac surgery. The LCP caused an ‘arc-shaped’ artefact (in the cine images because of local field distortion leading to de-phasing of the transverse magnetization). However, these artefacts impaired the diagnostic quality of the cardiac MR images only in a small region of the apex (Panel s B, C , and D). During and after the scan, no device related adverse events occurred. The LCP’s functional parameters were stable (pacing threshold 0.5 V and 0.38 V, impedance 550 Ω and 580 Ω, sensing 20 mV and 20 mV before and immediately after the scan, respectively). Support by the Austrian Research Promotion Agency FFG, within the scope of project 853390 LaMiCellPro, is gratefully acknowledged. Supplementary material is available at European Heart Journal online.
No takes yet. Share an insight, caveat, or question.
Kypta et al. (2017) conducted a case report in Leadless cardiac pacemaker (n=1). 3.0 Tesla cardiac magnetic resonance imaging was evaluated on Device related adverse events and functional parameters. 3.0 Tesla cardiac MRI was safely performed in a patient with a leadless cardiac pacemaker, with no device-related adverse events and stable functional parameters.