Answers are grounded in this paper and cite their sources.
Nuclear medicine imaging devices commonly use multi-element photo detection systems, composed of an array of N × N elements, each one providing an individual signal. Many strategies have been developed to reduce the number of readout channels, one of the main approaches is the Rows and Columns (R/C) projection logic. In this paper we proposed a modified version of Raised To the Power (RTP) algorithm adapted to R/C logic. In order to validate its efficiency a linear scanning irradiation on two 49 × 49 mm 2 LaBr 3 :Ce (0.5%) crystals with different thickness (4 mm and 10 mm) was carried out. Imaging performance analysis was made in terms of position linearity, Field-of-View (FoV) enlargement and spatial resolution. Imaging results from Anger Logic, RTP algorithm based on single element readout and RTP algorithm based on R/C readout were compared. A notable advantage of using RTP algorithms instead of Anger Logic was found: the FoV widens from about 30% to more than 70% of the detector area whereas the spatial resolution is highly improved, especially for off-center interactions, both for 4 mm-thick and 10 mm-thick crystals. Furthermore, imaging performance with the R/C readout is just slightly different from the single element one (FoV reduction less than 7% and SR worsening less than 10%). The R/C adapted RTP algorithm opens doors to high imaging performance with a substantial reduction of complexity and cost in the readout electronics.
No takes yet. Share an insight, caveat, or question.
Pani et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: