Key result
AliveCor 1L ECG algorithm detects AF with only ~51% sensitivity vs cardiologist overread.
Why the study?
The real-world performance of handheld single-lead ECG devices for atrial fibrillation screening in primary care is not well understood.
Does the AliveCor KardiaMobile 1L ECG automated algorithm accurately detect atrial fibrillation compared to cardiologist overread in individuals aged ≥65 years in primary care?
Cross-Sectional (n=14,230)
Unblinded
Does the AliveCor KardiaMobile 1L ECG automated algorithm accurately detect atrial fibrillation compared to cardiologist overread in individuals aged ≥65 years in primary care?
Automated interpretations of handheld single-lead ECGs in older primary care patients are sufficiently accurate to exclude AF but require physician overread to establish a diagnosis due to low positive predictive value.
Requires cardiologist overread for 1L ECG AF screening; leaves open performance in unselected real-world populations.
Handheld single-lead electrocardiographic (1L ECG) devices are increasingly used for atrial fibrillation (AF) screening, but their real-world performance is not well understood. The purpose of this study was to quantify the diagnostic test characteristics of 1L ECG automated interpretations for prospective AF screening. We calculated the diagnostic test characteristics of the AliveCor KardiaMobile 1L ECG (AliveCor, US) algorithm using unblinded cardiologist overread as the gold standard using single 30s tracings administered by medical assistants among individuals aged ≥65 years participating in the VITAL-AF trial (NCT03515057) of population-based AF screening embedded within routine primary care. A total of 14,230 individuals (mean age 74 ± 7 years, 60% women, 82% White) had 31,376 tracings reviewed by 13 cardiologists. A total of 24,906 (79.6%) tracings had an AliveCor interpretation of normal, 5,046 (16.1%) were unclassified, 797 (2.5%) were possible AF, and 573 (1.8%) were no analysis. Cardiologists read 808 (2.6%) tracings as AF. AliveCor possible AF had a PPV of 51.7% (95% CI: 47.8%-55.6%). AliveCor normal had an NPV of 99.8% (95% CI: 99.7%-99.8%). The AliveCor algorithm had an overall sensitivity of 51.0% (95% CI: 47.1%-54.9%) and a specificity of 98.7% (95% CI: 98.6%-98.9%). AliveCor tracings interpreted as unclassified (PPV 5.9%, 95% CI: 5.1%-6.7%) and no analysis (PPV 6.5%, 95% CI: 4.6%-8.9%) had low predictive values for AF and were increasingly prevalent at older ages (13.7% for age 65-69 years to 28.1% for age ≥85 years, P < 0.01). In an older primary care population undergoing AF screening with handheld 1L ECGs, automated algorithm interpretations were sufficiently accurate to exclude the presence of AF but not to establish an AF diagnosis.
No takes yet. Share an insight, caveat, or question.
Khurshid et al. (2023) conducted a cross-sectional in Atrial fibrillation (n=14,230). AliveCor KardiaMobile 1L ECG automated interpretation vs. Unblinded cardiologist overread was evaluated on Diagnostic sensitivity for atrial fibrillation (95% CI 47.1-54.9). The AliveCor 1L ECG automated algorithm had a sensitivity of 51.0% (95% CI 47.1-54.9%) and specificity of 98.7% for detecting atrial fibrillation compared to cardiologist overread.