Key result
Intracavitary electrocardiography discriminates SVC from CAJ locations during cannula positioning with an AUC of 0.99.
Why the study?
Does intracavitary electrocardiography accurately localize percutaneous venous drainage cannula tips compared to transesophageal echocardiography in adult patients undergoing elective cardiac surgery?
Observational (n=20)
No
Does intracavitary electrocardiography accurately localize percutaneous venous drainage cannula tips compared to transesophageal echocardiography in adult patients undergoing elective cardiac surgery?
Effect estimate: AUC 0.99 (95% CI 0.88-1.00)
p-value: p=<0.0001
Intracavitary electrocardiography provides a highly accurate, safe, and feasible alternative to transesophageal echocardiography for real-time localization of large-bore percutaneous venous drainage cannulas during cardiopulmonary bypass.
May support IC-ECG as TEE alternative for SVC cannula positioning in CPB; hypothesis-generating, needs prospective validation before practice change.
Introduction: Accurate positioning of percutaneous venous drainage cannulas in the superior vena cava (SVC) is essential to ensure effective venous return during cardiopulmonary bypass (CPB), particularly in minimally invasive and re-operative cardiac surgery. Transesophageal echocardiography (TEE) is the reference standard for cannula tip localization but may not always be available or feasible. Intracavitary electrocardiography (IC-ECG) is an established technique for central venous catheter tip localization; however, its application to large-bore venous drainage cannulas used for CPB has not been previously investigated. Methods: This prospective single-center observational study enrolled 20 adult patients undergoing elective cardiac surgery requiring percutaneous SVC cannulation via the right internal jugular vein. IC-ECG signals were acquired in real-time through the cannula–introducer assembly during cannula advancement. Cannula tip position was simultaneously assessed using TEE as the reference standard. Mean P-wave amplitudes were measured at predefined anatomical locations: SVC, cavo-atrial junction (CAJ), and right atrium (RA), when feasible. Receiver operating characteristic (ROC) curve analysis was used to evaluate the diagnostic accuracy of mean P-wave amplitude in identifying correct cannula tip positioning at the CAJ, compared with SVC and RA. Results: Interpretable IC-ECG signals were obtained in all patients. No adverse events were observed. Mean P-wave amplitude was significantly higher at the CAJ compared with the SVC and RA (7.4 vs 2.6 and 5.2 mm, respectively; p = 0.02). ROC analysis demonstrated effective discrimination between SCV and CAJ locations with an area under the curve of 0.99 (95% CI: 0.88–1.00; p < 0.0001). In the group of 15 patients in whom advancement into RA was feasible, IC-ECG demonstrated high diagnostic accuracy in discriminating between CAJ and RA location. Conclusions: IC-ECG is a feasible, safe, and accurate method for real-time localization of large-bore percutaneous venous drainage cannula tips, expanding its applicability within the field of vascular access.
No takes yet. Share an insight, caveat, or question.
Calabrese et al. (2026) conducted an observational in Elective cardiac surgery requiring percutaneous SVC cannulation (n=20). Intracavitary electrocardiography (IC-ECG) vs. Transesophageal echocardiography (TEE) was evaluated on Diagnostic accuracy of mean P-wave amplitude in identifying correct cannula tip positioning at the cavo-atrial junction (AUC 0.99, 95% CI 0.88-1.00, p=<0.0001). Intracavitary electrocardiography demonstrated high diagnostic accuracy in discriminating between SVC and CAJ locations for cannula tip positioning (AUC 0.99; 95% CI 0.88-1.00; p<0.0001).