Key result
Wearable carotid ultrasound flow time correlates strongly with velocity time integral during central hypovolemia.
Why the study?
Conflicting data exist regarding the correlation between common carotid artery corrected flow time and velocity time integral as surrogates of changing stroke volume, possibly due to inadequate beat sample size and measurement variability.
Does common carotid artery corrected flow time (ccFT) correlate with velocity time integral (VTI) during central blood volume loss and resuscitation in healthy adults?
Observational (n=16)
Does common carotid artery corrected flow time (ccFT) correlate with velocity time integral (VTI) during central blood volume loss and resuscitation in healthy adults?
Effect estimate: R = 0.81 (10 cycles)
There is a strong linear correlation between carotid artery corrected flow time and velocity time integral during central blood volume changes, supporting its potential use as a surrogate for central hemodynamics.
Strong ccFT-VTI correlation in healthy volunteers with averaged cycles; leaves open clinical validation as stroke volume surrogate.
Background: Doppler ultrasound of the common carotid artery is used to infer central hemodynamics. For example, change in the common carotid artery corrected flow time (ccFT) and velocity time integral (VTI) are proposed surrogates of changing stroke volume. However, conflicting data exist which may be due to inadequate beat sample size and measurement variability – both intrinsic to handheld systems. In this brief communication, we determined the correlation between changing ccFT and carotid VTI during progressively severe central blood volume loss and resuscitation. Methods: Measurements were obtained through a novel, wireless, wearable Doppler ultrasound system. Sixteen participants (ages of 18–40 years with no previous medical history) were studied across 25 lower body-negative pressure protocols. Relationships were assessed using repeated-measures correlation regression models. Results: In total, 33,110 cardiac cycles comprise this analysis; repeated-measures correlation showed a strong, linear relationship between ccFT and VTI. The strength of the ccFT-VTI relationship was dependent on the number of consecutively averaged cardiac cycles (R 1 cycle = 0.70, R 2 cycles = 0.74, and R 10 cycles = 0.81). Conclusions: These results positively support future clinical investigations employing common carotid artery Doppler as a surrogate for central hemodynamics.
No takes yet. Share an insight, caveat, or question.
Kerrebijn et al. (2023) conducted an observational in Healthy (no previous medical history) (n=16). Lower body-negative pressure protocols was evaluated on Correlation between changing common carotid artery corrected flow time (ccFT) and velocity time integral (VTI) (R = 0.81 (10 cycles)). Common carotid artery corrected flow time and velocity time integral showed a strong linear relationship during central blood volume loss (R=0.81 for 10 consecutively averaged cardiac cycles).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: