Key result
Dual-microphone narrowband filtering reduces acoustic noise and enhances blood pressure pulse signals.
Why the study?
Current auscultatory blood pressure measurement is unreliable in high noise situations such as ambulatory environments or moderate patient movement.
A novel dual-microphone technique for auscultatory blood pressure measurement can theoretically filter out acoustic noise, potentially improving reliability in ambulatory or high-noise environments.
Answers are grounded in this paper and cite their sources.
Dual-mic narrowband filtering may enable auscultatory BP amid noise; leaves open prospective validation before clinical use.
Auscultatory blood pressure measurement uses the presence and absence of acoustic pulses generated by an artery (i.e., Korotkoff sound), detected with a stethoscope or a sensitive microphone, to noninvasively estimate systolic and diastolic pressures. Unfortunately, in high noise situations, such as ambulatory environments or when the patient moves moderately, the current auscultatory blood pressure method is unreliable, if at all possible. Empirical evidence suggests that the pulse beneath an artery occlusion travels relatively slow compared with the speed of sound. By placing two microphones along the bicep muscle near the brachial artery under the occlusion cuff, a similar blood pressure pulse appears in the two microphones with a relative time delay. The acoustic noise, on the other hand, appears in both microphones simultaneously. The contribution of this paper is to utilize this phenomenon by filtering the microphone waveforms to create spatially narrowband information signals. With a narrowband signal, the microphone signal phasing information is adequate for distinguishing between acoustic noise and the blood pressure pulse. By choosing the microphone spacing correctly, subtraction of the two signals will enhance the information signal and cancel the noise signal. The general spacing problem is also presented.
No takes yet. Share an insight, caveat, or question.
Sebald et al. (2002) studied Blood pressure measurement. Narrowband auscultatory blood pressure measurement using two microphones vs. Standard auscultatory blood pressure method was evaluated on Distinguishing between acoustic noise and blood pressure pulse. Using two microphones and filtering waveforms to create spatially narrowband signals allows for the subtraction of acoustic noise and enhancement of the blood pressure pulse signal.
Synapse has enriched 10 related papers. Consider them for comparative context: