A flow injection analysis‐resonance light scattering (FIA‐RLS) coupled technique was developed originally to determine hepanrin in the presence of Azure B (AB). The FIA‐RLS coupled technique overcomes the drawbacks of RLS technique such as poor reproducibility and tedious operation without sacrificing sensitivity and simplicity. The method is based on the enhanced weak RLS intensity of AB by the addition of a trace of heparin (HP) with λmax located at 340.0 nm at pH 7.96. In time scanning mode, a linear relationship between FIA‐RLS intensity and heparin concentration in the range of 0.01–6.0 µg ml−1 was established with the limit of detection (LOD) of 6.4 ng ml−1 (3σ) with a relative standard deviation (RSD) of 2.1% (n = 20, S/B > 3), and the LOQ of 10 ng ml−1. Owing to the introduction of FIA automation, the throughput of FIA‐RLS coupled technique could be more than 120 samples per hour.