Key result
Migrated LAmbre LAA occlusion device causes mitral entrapment, requiring emergency surgical valve replacement.
Why the study?
Although left atrial appendage occlusion is generally safe, life-threatening complications such as device embolization and cardiac tamponade can occur.
Case Report (n=1)
Highlights a life-threatening complication of LAAO device migration requiring emergency surgery and resulting in mitral valve replacement.
Answers are grounded in this paper and cite their sources.
Delayed LAmbre migration can trigger emergency surgery; this Level 5 case leaves open true incidence and risk factors.
Left atrial appendage occlusion has become an alternative for long-term anticoagulation for patients with non-valvular atrial fibrillation. Although the procedure is safe, life-threatening complications such as embolization of the device or cardiac tamponade might occur. We present a case of a LAmbre device that migrated 4 days after being implanted and remained trapped in the mitral valve. Secondary massive mitral regurgitation with severe stenosis and haemodynamic instability required emergency surgery. The device was successfully removed, but severe damage in the anterior leaflet and chords forced a valve replacement.
No takes yet. Share an insight, caveat, or question.
Martínez-López et al. (2021) conducted a case report in Non-valvular atrial fibrillation (n=1). LAmbre device implantation was evaluated on Device migration and clinical consequences. A LAmbre left atrial appendage occlusion device migrated 4 days post-implantation and became trapped in the mitral valve, requiring emergency surgical removal and valve replacement.