Key result
WATCHMAN hook protrusion leads to pulmonary artery perforation and pericardial tamponade.
Why the study?
Pulmonary artery perforation caused by the WATCHMAN left atrial appendage closure device had not been previously reported, highlighting a need for awareness of adjacent structure risks.
Case Report (n=1)
This is the first reported case of pulmonary artery perforation caused by WATCHMAN device barbs, emphasizing the importance of evaluating structures adjacent to the LAA prior to closure.
Answers are grounded in this paper and cite their sources.
Alerts to rare hook-related perforation risk with WATCHMAN; leaves open need for safety data in larger cohorts.
We describe a female patient who received hybrid catheter intervention and occlusion of the patent ductus arteriosus (PDA) and left atrial appendage (LAA). Four hours after the procedure, pericardial tamponade suddenly occurred. Surgical exploration of the heart found that a fixation hook of the WATCHMAN device, had protruded through the LAA wall and tore the dilated pulmonary artery (PA). This is the first case report to describe PA perforation caused by barbs of the WATCHMAN device, and more attention should be paid to the relationship between the LAA and its adjacent structures before LAA closure.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2019) conducted a case report in Patent ductus arteriosus and left atrial appendage closure (n=1). WATCHMAN device was evaluated on Pulmonary artery perforation and pericardial tamponade. Left atrial appendage closure using the WATCHMAN device caused pulmonary artery perforation and pericardial tamponade due to a fixation hook protruding through the LAA wall.
Synapse has enriched 10 related papers. Consider them for comparative context: