Answers are grounded in this paper and cite their sources.
Randomized trial demonstrates that patient shielding drapes reduce occupational radiation dose to interventional cardiologists by half, indicating effective staff protection.
The aim of this study was to evaluate the occupational radiation dose in interventional cardiology by using a shielding drape on the patient. A random study with and without the protective material was conducted. The following control parameters were registered: demographic data, number of stents, contrast media volume, fluoroscopy time, number of cine images, kerma-area product and cumulative air kerma. Occupational dose data were obtained by electronic active dosemeters. No statistically significant differences in the analysed control parameters were registered. The median dose value received by the interventional cardiologist was 50% lower in the group with a shielding drape with a statistically significant p-value <0.001. In addition, the median value of the maximum scatter radiation dose was 31% lower in this group with a statistically significant p-value <0.001. This study showed that a shielding drape is a useful tool for reducing the occupational radiation dose in a cardiac catheterisation laboratory.
No takes yet. Share an insight, caveat, or question.
Ordiales et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: