Key result
Real-time PC-MRI matches gated PC-MRI for lumen area and stroke volume within ~5%.
Why the study?
Accurate and reproducible measurements of peripheral artery blood flow and vessel area variations are needed to study their correlation with cardiovascular risk factors.
Answers are grounded in this paper and cite their sources.
The developed real-time PC-MRI method provides accurate measurements of blood flow and lumen area variations in peripheral arteries.
PURPOSE: To measure peripheral artery function using a real-time phase-contrast (PC)-MRI sequence with tailored image-processing algorithms for flow computation. MATERIALS AND METHODS: An approach to real-time flow measurements was developed based on two-dimensional spatially selective excitation pulses and consecutive tailored processing of the data to derive blood flow and vessel area variations. The data acquisition strategy allows for flow measurements at high spatial and temporal resolutions of 1 mm(2) and 50 msec, respectively. In postprocessing the vessel area is automatically extracted using correlation measures in conjunction with morphological image operators. By means of in vitro and in vivo validations, it is shown that the current methods provide accurate and reproducible measurements of flow and vessel area variations. RESULTS: In vitro the comparison between the lumen area measured with the presented method and the values obtained by caliper gauge measurement showed a difference of 3.4% +/- 3.4% (mean +/- 2 SD). Similarly, the comparison between the stroke volumes determined with the presented method and by stopwatch and bucket measurements yielded a difference of 6.1% +/- 2.1%. In vivo the results from the real-time measurements for lumen area and stroke volume were compared with those from a gated PC-MRI technique with differences of 4.8% +/- 14% and 3.0% +/- 24.7%, respectively. CONCLUSION: The presented method constitutes a reliable tool set for quantifying the variations of blood flow and lumen area in the superficial femoral artery during reactive hyperemia and for studying their correlation with cardiovascular risk factors.
No takes yet. Share an insight, caveat, or question.
Oelhafen et al. (2006) studied Peripheral artery function. Real-time phase-contrast (PC)-MRI vs. Gated PC-MRI (in vivo) and manual measurements (in vitro) was evaluated on Difference in lumen area and stroke volume. Real-time phase-contrast MRI provided accurate measurements of lumen area and stroke volume in vivo, with differences of 4.8% ± 14% and 3.0% ± 24.7% compared to gated PC-MRI.
Synapse has enriched 10 related papers. Consider them for comparative context: