Key result
Auscultatory and oscillometric BP norms are interchangeable for pediatric SBP, but not for DBP.
Why the study?
It was unclear whether auscultatory normative values can be applied to blood pressure obtained by oscillometric devices in children.
Can auscultatory blood pressure normative values be used for evaluation of oscillometric blood pressure in children?
Observational (n=229)
Can auscultatory blood pressure normative values be used for evaluation of oscillometric blood pressure in children?
Auscultatory and oscillometric normative data are interchangeable for systolic blood pressure but not for diastolic blood pressure in children.
Answers are grounded in this paper and cite their sources.
May support auscultatory norms for oscillometric SBP in children; leaves open DBP accuracy pending prospective validation.
The aim of the study was to analyze whether auscultatory normative values (Fourth Task Force [4TF]) can be applied to blood pressure (BP) obtained by oscillometric devices. The authors performed a retrospective analysis of oscillometric office BP and ambulatory BP monitoring in 229 children (116 boys), median age 15.31 years. Office systolic BP (SBP) and diastolic BP (DBP) values were converted into Z scores using 4TF and oscillometric (German Health Interview and Examination Survey for Children and Adolescent [KiGGS]) reference values. There was good correlation between the two normative methods (r=0.9773 for SBP, r=0.9627 for DBP). Results from Bland-Altman test revealed only minimal differences in Z scores between 4TF and KiGGS for SBP, but a significant proportional error for DBP. 4TF and KiGGS Z scores were equally predictive of ambulatory hypertension. In conclusion, auscultatory and oscillometric normative data are interchangeable for SBP but not for DBP.
No takes yet. Share an insight, caveat, or question.
Šuláková et al. (2016) conducted an observational in Blood pressure evaluation (n=229). Auscultatory normative values (4TF) vs. Oscillometric normative values (KiGGS) was evaluated on Correlation and agreement of Z scores for SBP and DBP between 4TF and KiGGS normative values (r=0.9773 for SBP, r=0.9627 for DBP). Auscultatory normative values (4TF) and oscillometric reference values (KiGGS) showed good correlation and are interchangeable for SBP but not for DBP.