Why the study?
Complex vascular anatomy may challenge the deployment of the dual-filter Sentinel Cerebral Protection System during TAVR, but the impact of specific anatomic features on device failure was unclear.
Do anatomic features of the aortic arch and supra-aortic arteries predict technical device failure of dual-filter cerebral embolic protection in patients undergoing TAVR?
Observational (n=92)
Do anatomic features of the aortic arch and supra-aortic arteries predict technical device failure of dual-filter cerebral embolic protection in patients undergoing TAVR?
Answers are grounded in this paper and cite their sources.
Brachiocephalic tortuosity and an angle >59° are associated with technical failure of dual-filter cerebral embolic protection during TAVR, suggesting this anatomy should preclude its use.
BACKGROUND: The dual-filter Sentinel™ Cerebral Protection System (Sentinel-CPS) is increasingly used during transcatheter aortic valve replacement (TAVR). However, complex vascular anatomy may challenge Sentinel-CPS deployment. AIM OF THE STUDY: We sought to investigate the impact of anatomic features of the aortic arch and the supra-aortic arteries on technical device failure of Sentinel-CPS application. METHODS: Analysis of the multislice computed tomography pre-TAVR aortograms of all patients undergoing TAVR with Sentinel-CPS between 2016 and 2020 (n = 92) was performed. We investigated the impact of aortic arch anatomy, configuration, and the angles of the supra-aortic arteries, including the determination of vascular tortuosity index on device failure of Sentinel-CPS application. RESULTS: The Sentinel-CPS was applied successfully in 83 patients (90.2%). Device failure in nine patients (9.8%) was due to the infeasibility to perform correct deployment of both filters (n = 7) and to obtain peripheral radial access (n = 2). Patients with a failure of Sentinel-CPS application had a higher right subclavian tortuosity index (217 [92-324] vs. 150 [42-252], p = .046), a higher brachiocephalic tortuosity index (27 [5-51] vs. 10 [0-102], p = 0.033) and a larger angulation of the brachiocephalic artery (59° [22-80] vs. 39° [7-104], p = .014) compared with patients with successful application. A brachiocephalic angle more than 59° was predictive for device failure. No differences in aortic arch anatomy or common carotid artery tortuosity were detected between the groups. CONCLUSIONS: Brachiocephalic tortuosity was found to be associated with failure of Sentinel-CPS application. Filter-based usage should be avoided in TAVR patients with a brachiocephalic angle more than 59°.
No takes yet. Share an insight, caveat, or question.
Campanella et al. (2021) conducted an observational in Transcatheter aortic valve replacement (TAVR) (n=92). Complex vascular anatomy (brachiocephalic tortuosity and angulation) vs. Less complex vascular anatomy was evaluated on Technical device failure of Sentinel-CPS application.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: