Key result
Transcatheter PDA closure achieves ~98% procedural success with minimal mortality.
Why the study?
The efficacy, safety, and immediate complications of percutaneous device closure of patent ductus arteriosus needed to be determined.
Does percutaneous device closure safely and effectively treat patent ductus arteriosus?
Observational (n=500)
No
Does percutaneous device closure safely and effectively treat patent ductus arteriosus?
Transcatheter occlusion of PDA is an effective and safe therapeutic option with a high success rate (98.2%) and low complication rate.
Answers are grounded in this paper and cite their sources.
Large case series supports procedural feasibility for PDA closure; leaves open need for controlled comparisons on device choice and durability.
OBJECTIVE: To determine the efficacy, safety and immediate complications encountered during percutaneous device closure of patent ductus arteriosus (PDA). STUDY DESIGN: Case series. PLACE AND DURATION OF STUDY: Department of Paediatric Cardiology, AFIC/NIHD, Rawalpindi, from January 2005 to December 2010. METHODOLOGY: Consecutive 500 patients who underwent attempted transcatheter PDA device closure were included in the study. Device type position, success of closure and complications were described as frequency percentage. RESULTS: In 491 cases (98.2%), PDA was successfully occluded including 4 cases (0.8%) where devices were dislodged but retrieved and redeployed in Cath laboratory. PDA occluder devices used in 448 cases (91%) while coils (single or multiple) were used in 42 cases (8.5%) and in one case (0.2%) ASD occluder device was used to occlude the PDA. There were 09 (1.8%) unsuccessful cases, 06 (1.2%) were abandoned as ducts were considered unsuitable for device closure, 02 (0.4%) devices dislodged and needed surgical retrieval and one case (0.2%) was abandoned due to faulty equipment. The narrowest PDA diameter ranged from 0.5-14 mm with mean of 4.5±2.4 mm. There was a single (0.2%) mortality. CONCLUSION: Transcatheter occlusion of PDA by coil or occluder device is an effective therapeutic option with high success rate. Complication rate is low in the hands of skilled operators yet paediatric cardiac surgical back-up cover is mandatory.
No takes yet. Share an insight, caveat, or question.
Sultan et al. (2014) conducted an observational in Patent ductus arteriosus (n=500). Transcatheter PDA device closure was evaluated on Successful occlusion of PDA. Transcatheter device closure of patent ductus arteriosus was successfully achieved in 98.2% of cases, with a low mortality rate of 0.2%.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: