Answers are grounded in this paper and cite their sources.
This paper introduces a 6.78-MHz wireless power transfer (WPT) system for implantable medical devices (IMDs). The proposed dual-output resonant current-mode (RCM) rectifier cumulates energy from loosely-coupled coils and generates two output voltages (VH = 3-V, VL = 1.8-V) regulated by adaptive power control (APC) and local-loop power control. Due to transmission power regulation (TPR), the transmitter (TX) delivers appropriate power to the receiver (RX) to realize global-loop power control. Thus, the power transfer efficiency (PTE) is improved, especially under light load. Furthermore, zero-voltage switching and zero-current switching techniques enhanceR Xpower conversion efficiency (PCE).TXandRXchips were fabricated in a 0.18-μ mCMOS process. The measurement results show that the proposed WPT system successfully regulates outputs at VH = 3-V and VL = 1.8-V at a 1.5-cm coupling distance. With the proposed TPR, PTE is improved by 28.2% at PTOTAL = 1.6-mW, and input power is reduced by 94.8% at PTOTAL = 1-mW. The measured peak PCE and peak PTE are 85.1% and 31.3% at a coil distance of 10-mm, respectively.
No takes yet. Share an insight, caveat, or question.
Yao et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: