Key result
Suspended Tesla pump achieves 77 mm Hg head rise at target flow with antithrombotic shear.
Why the study?
Computational fluid dynamics was used to design and analyze a novel suspended Tesla left ventricular assist device to understand parameters affecting device performance.
Computational fluid dynamics modeling of a novel suspended Tesla LVAD demonstrated favorable hydrodynamic performance and wall shear stresses, suggesting potential for further development.
Supports further preclinical LVAD development; leaves open clinical translation of the Tesla pump design.
This article summarizes the use of computational fluid dynamics (CFD) to design a novel suspended Tesla left ventricular assist device. Several design variants were analyzed to study the parameters affecting device performance. CFD was performed at pump speeds of 6500, 6750, and 7000 rpm and at flow rates varying from 3 to 7 liters per minute (LPM). The CFD showed that shortening the plates nearest the pump inlet reduced the separations formed beneath the upper plate leading edges and provided a more uniform flow distribution through the rotor gaps, both of which positively affected the device hydrodynamic performance. The final pump design was found to produce a head rise of 77 mm Hg with a hydraulic efficiency of 16% at the design conditions of 6 LPM through flow and a 6750 rpm rotation rate. To assess the device hemodynamics the strain rate fields were evaluated. The wall shear stresses demonstrated that the pump wall shear stresses were likely adequate to inhibit thrombus deposition. Finally, an integrated field hemolysis model was applied to the CFD results to assess the effects of design variation and operating conditions on the device hemolytic performance.
No takes yet. Share an insight, caveat, or question.
Medvitz et al. (2010) studied this question. Suspended Tesla left ventricular assist device was evaluated on Device hydrodynamic performance (head rise and hydraulic efficiency). The final suspended Tesla pump design produced a head rise of 77 mm Hg with a hydraulic efficiency of 16% at 6 LPM flow and 6750 rpm, with wall shear stresses likely adequate to inhibit thrombosis.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: