Key result
An LC-MS/MS assay was successfully developed and validated for the combined therapeutic drug monitoring of five PARP inhibitors, meeting FDA and EMA guidelines.
A simple and fast LC-MS/MS assay was successfully developed and validated for the therapeutic drug monitoring of five PARP inhibitors.
Answers are grounded in this paper and cite their sources.
Provides validated tool for PARP inhibitor TDM; clinical utility and outcome impact remain to be established.
An liquid chromatography-mass spectrometry (LC-MS/MS) assay was developed for the combined analysis of the five poly (ADP-ribose) polymerase (PARP) inhibitors niraparib, olaparib, rucaparib talazoparib and veliparib. A simple and fast sample pre-treatment method was used by protein precipitating of plasma samples with acetonitrile and dilution of the supernatant with formic acid (0.1% v/v in water). This was followed by chromatographic separation on a reversed-phase UPLC BEH C18 column and detection with a triple quadrupole mass spectrometer operating in the positive mode. A simplified validation procedure specifically designed for bioanalytical methods for clinical therapeutic drug monitoring (TDM) purposes, was applied. This included assessment of the calibration model, accuracy and precision, lower limit of quantification (LLOQ), specificity and selectivity, carry-over and stability. The validated range was 30-3000 ng/mL for niraparib, 100-10,000 ng/mL for olaparib, 50-5000 ng/mL for rucaparib, 0.5-50 ng/mL for talazoparib and 50-5000 for veliparib. All results were within the criteria of the US Food and Drug Administration (FDA) guidance and European Medicines Agency (EMA) guidelines on method validation. The assay has been successfully implemented in our laboratory.
No takes yet. Share an insight, caveat, or question.
Bruin et al. (2019) studied this question. LC-MS/MS assay for PARP inhibitors was evaluated on Assay validation parameters (calibration, accuracy, precision, LLOQ, specificity, selectivity, carry-over, stability). An LC-MS/MS assay was successfully developed and validated for the combined therapeutic drug monitoring of five PARP inhibitors, meeting FDA and EMA guidelines.
Synapse has enriched 10 related papers. Consider them for comparative context: