Key result
Synchronized biphasic defibrillation fails to improve shock success over non-synchronized defibrillation for VF.
Why the study?
Optimization of defibrillation success is important to improve efficacy and minimize post-shock sequelae, and previous work suggested improved shock success with intracardiac shocks synchronized to the upslope of a VF wave.
Does synchronized defibrillation improve shock success compared to non-synchronized defibrillation in a porcine model of ventricular fibrillation?
RCT (n=10)
Randomized
No
Does synchronized defibrillation improve shock success compared to non-synchronized defibrillation in a porcine model of ventricular fibrillation?
Synchronized defibrillation to the upslope of the surface ECG did not improve shock success compared to non-synchronized shocks in a porcine model of ventricular fibrillation.
Answers are grounded in this paper and cite their sources.
No benefit from synchronization in porcine VF; leaves open whether timing matters in human defibrillation.
OBJECTIVE: Optimization of defibrillation success is important to improve efficacy and minimize post-shock sequelae. Previous work has suggested an improvement in shock success when an intracardiac shock is delivered synchronized to the upslope of a VF wave. We investigated the efficacy of transthoracic defibrillation success using a novel external biphasic defibrillator which delivers shocks synchronized to the upslope of the surface ECG. METHODS: A prospective, controlled, randomized study in a research institute laboratory of male and female pigs (54.2±1.8 kg). Ventricular fibrillation (VF) was induced in 10 anaesthetized and ventilated pigs. Shocks were delivered randomly from a biphasic defibrillator in synchronized or non-synchronized mode via self-adhesive electrode pads following 30 s of VF. Energy settings at 50, 70, 80, and 100J were randomly tested. VF amplitude, impedance, and shock outcome were recorded and analysed digitally. RESULTS: A total of 300 shocks were delivered. Synchronized shocks were delivered on the upslope of the VF wave in 99% of cases. There was no significant difference in shock success between shocks delivered in synchronized or non-synchronized modes (p=0.695). There was no significant difference in the amplitude of VF between successful and unsuccessful shocks (p=0.163). Furthermore, there was no association between shock success and transthoracic impedance. CONCLUSION: The novel defibrillator used in this study was able to consistently deliver shocks on the upslope portion of the VF wave but did not show an improvement in shock success.
No takes yet. Share an insight, caveat, or question.
Darragh et al. (2012) conducted an RCT in Ventricular fibrillation (n=10). Synchronized biphasic defibrillation vs. Non-synchronized biphasic defibrillation was evaluated on Shock success (p=0.695). Synchronized biphasic defibrillation did not improve shock success compared to non-synchronized defibrillation for ventricular fibrillation.