During emergency transfusion, stored whole blood at 1-6°C must be rapidly warmed to 36-38°C using blood warming devices. However, device validation commonly relies on Newtonian crystalloid intravenous fluids with a viscosity of 1 mPa·s, whereas whole blood exhibits temperature-dependent, shear-thinning rheology with viscosities of 3.5-4.5 mPa·s at 37°C. This discrepancy introduces systematic errors in performance evaluation. To address this gap, this research study characterized three blood-mimicking fluids (BMFs), including glycerol-water solutions of 22-50% w/w, egg whites, and TrueClot across temperatures of 5-40°C and shear rates of 5-100 s⁻¹. Among the tested fluids, TrueClot most closely resembled blood in overall rheological properties. However, TrueClot demonstrated abnormal phase separation at temperatures below 10°C, with a 28% drop in viscosity from 55 ± 4 to 39.6 ± 0.5 mPa·s over just 2°C, limiting its use at refrigerated storage temperatures. At shear rates ≥ 100 s⁻¹, 50% glycerol-water solutions produced viscosity values closest to reported whole blood values. Based on the rheological profiles obtained, it is inferred that at shear rates < 100 s⁻¹, diluted TrueClot at temperatures higher than 10°C or modified egg whites with xanthan gum at 5-40°C may provide suitable shear-thinning properties, though experimental validation of these modified solutions is required. Two mathematical models, including the Casson model with Apostolidis and Beris parametrization and a combined shear-temperature power law model, were developed to predict viscosity values that correlate well with the experimental results.
No takes yet. Share an insight, caveat, or question.
Riddle et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: