Double transeptal technique enables SAPIEN M3 TMVR in challenging anatomies like small LVOT.
Why the study?
Certain clinical scenarios such as low dock position or small LVOT may require a second guide sheath deployment during SAPIEN M3 TMVR, but experience with double transeptal technique is limited.
Case Report (n=2)
Answers are grounded in this paper and cite their sources.
Describes a double transeptal technique for SAPIEN M3 TMVR to address challenging clinical scenarios like low dock position or small LVOT.
The SAPIEN M3 system has been recently approved for transcatheter mitral valve replacement (TMVR). The transeptal guide sheath is used to deploy the dock, which is completely released before inserting and deploying the M3 valve using the same femoral access, transeptal puncture (TSP) and guide sheath. However, clinical scenarios such us low dock position or small left ventricular outflow tract (LVOT) may require to deploy the M3 valve with a second guide sheath while the dock is still attached to the first. In this article, we describe our initial experience with the “double transeptal technique” in two patients, focusing on indications and key technical points.
No takes yet. Share an insight, caveat, or question.
Fernández‐Cordón et al. (2026) conducted a case report in Transcatheter mitral valve replacement (n=2). Double transeptal technique for SAPIEN M3 TMVR was evaluated. The double transeptal technique for SAPIEN M3 transcatheter mitral valve replacement was described in two patients with clinical scenarios such as low dock position or small LVOT.