Key result
Amplatzer Vascular Plug successfully resolves immediate severe paravalvular regurgitation after Sapien M3 valve replacement.
Why the study?
Paravalvular leak is a known complication of transcatheter mitral valve replacement devices, and this report illustrates its percutaneous management following Sapien M3 transcatheter heart valve implantation.
Case Report (n=1)
No
Intraprocedural severe paravalvular regurgitation following Sapien M3 transcatheter mitral valve replacement can be feasibly and successfully managed using an Amplatzer Vascular Plug.
Answers are grounded in this paper and cite their sources.
Requires vigilant PVL surveillance after Sapien M3 implantation; leaves open optimal prevention and management strategies.
Background Transcatheter mitral valve replacement by the Sapien M3 transcatheter heart valve offers a therapeutic option for patients with severe mitral regurgitation that are neither eligible for surgical therapy nor transcatheter edge-to-edge repair. This therapy allows treatment of patients with insufficient mitral valve annulus support by calcification or prior surgical mitral valve replacement. The presented case illustrates how paravalvular leak, a complication previously reported with other mitral valve replacement devices, can be successfully managed. Case Summary A 59 year old patient symptomatic with severe eccentric mitral regurgitation Carpentier IIIb underwent transcatheter mitral valve replacement with the Sapien M3 system. Post implantation, a severe paravalvular regurgitation occurred and was successfully treated with an Amplatzer Vascular Plug at index procedure. Discussion This case shows an immediate severe paravalvular regurgitation post implantation of a Sapien M3 valve. In the context of literature, paravalvular regurgitation seems to be a frequent result of Sapien M3 therapy that could occur at late follow-up but as shown also intraprocedural. The management of Sapien M3 associated paravalvular regurgitation by Amplatzer Vascular Plug is feasible and successful.
No takes yet. Share an insight, caveat, or question.
Epe et al. (2026) conducted a case report in Severe eccentric mitral regurgitation (Carpentier IIIb) and subsequent severe paravalvular leak after transseptal transcatheter mitral valve replacement (n=1). Amplatzer Vascular Plug for Sapien M3-associated paravalvular leak was evaluated on Resolution of severe paravalvular regurgitation. Amplatzer Vascular Plug implantation during the index procedure successfully resolved immediate severe paravalvular regurgitation in 1 patient (age 59) after Sapien M3 valve replacement.