Key result
Color Doppler detects significant acute aortic regurgitation with ~88% sensitivity versus electromagnetic flow.
Why the study?
A quantitative noninvasive method to evaluate acute aortic regurgitation using color Doppler flow needed validation against electromagnetic flow measurements.
Answers are grounded in this paper and cite their sources.
Color Doppler may grade acute AR in dogs; leaves open human validation and clinical adoption.
Color Doppler flow studies were performed on ten anesthetized open-chest dogs. Acute aortic regurgitation was created in the dogs by a special valve-spreading catheter. The magnitude of valvular regurgitation was determined by aortic electromagnetic flow recordings of regurgitant fraction. Arbitrarily-designated grades of aortic regurgitation: mild (4%-10%), moderate (11%-30%), and severe ( greater than 30%) were assigned on the basis of electromagnetic flow. We attempted to obtain studies of varying degrees of AR in each animal. Mean regurgitant fraction for the three grades were 6.8 +/- 0.6% (n = 11), 22.0 +/- 2.4% (n = 7), and 40.4 +/- 2.5 (n = 20), respectively (each P less than 0.05). By color Doppler flow assessment, the ratio of regurgitant jet height to the left ventricular dimension at the junction of the left ventricular outflow tract and the aortic annulus (JH/LVOH) was measured in each study. AR was classified by Doppler as grade I (mild), 1%-24%; II (moderate), 25%-64%; and III (severe), greater than or equal to 65% jet height/left ventricular outflow tract height. Color Doppler flow correlated well with flowmeter assessment of regurgitant fraction. Color Doppler flow tests had a calculated sensitivity of 88%, specificity of 83%, and predictive value of 85% for significant (moderate + severe) aortic regurgitation. Our data support the concept that this method of color Doppler flow assessment provides a quantitative noninvasive evaluation of aortic regurgitation.
No takes yet. Share an insight, caveat, or question.
Reeves et al. (1990) studied Acute aortic regurgitation (n=10). Color Doppler flow assessment vs. Aortic electromagnetic flow recordings was evaluated on Detection of significant (moderate + severe) aortic regurgitation (Sensitivity 88%, specificity 83%, predictive value 85%). Color Doppler flow assessment demonstrated 88% sensitivity, 83% specificity, and 85% predictive value for detecting significant acute aortic regurgitation compared to electromagnetic flow.
Synapse has enriched 10 related papers. Consider them for comparative context: